Specifications Model
Section Thickness Step Size Z-Axis Deflection, measured at all speeds
and amplitudes with capacitance devices traceable to National Standards
Blade Oscillation Frequency at Calibration Blade Oscillation Amplitude at Calibration Total Travel of Bath Table Bath Table Rise & Fall Speed
Cutting Head Advance Speed when Slicing
Cutting Head Advance Speed Resolution Cutting Head Advance Speed Max (fast mode) Cutting Head Retraction Speed Blade Oscillation Frequency
(amplitude dependent) Frequency Step Size
Blade Oscillation Amplitude (Nominal) Blade Oscillation Amplitude Step Size: Machine Dimensions LxWxH
(With magnifier and cold light options but without scope)
Tissue Bath and Tissue Mount Power Requirements (selectable) Power Rating Fuse Rating
Weight, lift handles supplied Shipping Weight
Shipping Dimensions, in wooden crate
Tissue bath is demountable for easy cleaning. 115VAC 60Hz or 230VAC 50Hz tba
(115V) (230V)
43 Kg (excluding microscope) 58 Kg (excluding microscope) 475mm x 540mm x 840mm
Tissue bath is demountable for easy cleaning. 115VAC 60Hz or 230VAC 50Hz tba
(115V) (230V)
43 Kg (excluding microscope) 58 Kg (excluding microscope) 475mm x 540mm x 840mm
25 Hz
2.25mm 19 mm
0.85 mm/sec maximum
From a minimum of -1.00 mm/sec, through the zero point to a maximum of +1.00 mm/sec
0.01 mm/sec
+/- 4.0 mm/sec 4.0 mm/sec
50Hz to120Hz 5 Hz
0.5 mm to 2.5 mm 0.25 mm (nominal) 390 x 390 x 260mm
25 Hz
2.0mm 19 mm
0.85 mm/sec maximum
From a minimum of -1.00 mm/sec, through the zero point to a maximum of +1.00 mm/sec
0.01 mm/sec
+/- 4.0 mm/sec 4.0 mm/sec
50Hz to 100Hz 5 Hz
0.5 mm to 2.0 mm 0.25 mm (nominal) 390 x 390 x 260mm
7000smz-2 0.001 mm Sub-micron
5000mz-2 0.001 mm
1-2 microns
Integrally Mounted
Cold Light Source Model 7000-2-1
To facilitate the careful slicing operation, it is most important that the progress of the blade through the tissue be clearly observed. This observation is used to ensure control of
the speed of advance and of the oscillation of the blade. Two elements are required for clear observation, light and magnification.
Product Features
• Ability to direct the light where needed with two flexible fiber optic light guides.
• Maintains temperature stability of the tissue or surrounding a.c.s.f.
• Light Intensity is adjustable by a rotary dial knob.
• 20W/12V Halogen lamp • 2000 hour life approx.
• Vibrating Microtomes with Magnifier
• Fan and electronic power supply • Adjustable with potentiometer
• Twin 500mm fiber optic light guides
Integrally Mounted Magnifying Glass Model 7000-1-3
For general global slices a magnifying glass offering approximately 2x magnification is sufficient. However, if specific loci are under study then an inspection microscope is desirable. The binocular inspection microscope with 10x-40x magnification, a zoom of 1x-4x and a working distance of 80mm is ideal for the task.
Product Features
• 115mm Diameter • 305mm Focal Length
• 2x Magnification
Integrally Mounted Inspection Microscope Model 7000-1-1 (x20-x40) Model 7000-1-2 (x5 and x10)
Page 1 |
Page 2 |
Page 3 |
Page 4