AIR MONITORING
Measurement Performance Based on the calibration method described above, the
ULTRA.sens® highly accurate O3
Application Example can be used to perform gas analyses. Fig. 8 shows the measurement results for a gradual change in O3
concentration in the ppb range. This enables a measuring range of 0-1000 ppb with a resolution of approx. 5 ppb (Fig. 9). For long-term stable O3
(AZF) is recommended. This option allows long-term stable gas analyses to be performed over an unlimited period of time. Drift is completely compensated for (Fig. 10). There is no cross- sensitivity to other gases in the ambient air (water vapour, carbon dioxide, etc.). Other substances such as sulphur dioxide (SO2
), nitrogen oxides (NOx with the
ULTRA.sens® ), etc. can also be measured simultaneously , thus compensating for cross-sensitivities if necessary. The concentration
measurement is compensated for gas density via an integrated temperature and pressure sensor, so that this infl uence can also be neglected.
gas analyses, the integration of an AutoZeroFunction
One current example of the wide range of possible applications for this innovative gas measurement technology will be cited here. Plant growth (plant physiology) depends on many external factors such as light, humidity, temperature, CO2
concentration and also the O3 concentrations. These O3
in the ambient air. For this purpose, the respiratory activity of plants is analysed in special test facilities [7]. One stress factor is the O3
content, which is supplied to the plants in small concentrations can be measured and regulated very precisely using the
setup described above. Fig. 11 shows such a setup from the company Plant Invent (Estonia). The device allows parallel non-destructive measurements of stomatal conductance, photosynthesis and transpiration of multiple plants. Parallel measurements speed up the workfl ow while also providing more comparable gas exchange data. The cuvette conditions are precisely controlled, and various stimuli can be applied.
content
Fig. 11: Jyrkki MAX is a multi-cuvette stationary gas exchange device designed to measure bigger plants like cereals, ferns, tomatoes etc. (Plant Invent Estonia)
Fig. 8: Gradual change in O3 concentration using a voltage-controlled ozone generator (silent discharge) Summary
The
ULTRA.sens gas measurement module from Wi.Tec-Sensorik GmbH described above is characterised by its precise determination of ozone concentrations from trace levels (ppb) up to the industrial vol.% range. This is achieved through exact mathematical calibration in accordance with ISO 13964. Multi-channel measurement enables high dynamic ranges (ppb to vol.%), which can open up completely new areas of application. It is also possible to measure other gases (e.g. NO2
)
in addition to ozone in the same setup. This is helpful in plasma technology application engineering. The versatility of this gas measurement technology was demonstrated using an application example from plant physiology.
References
[1 ] Batakliev et al. : Ozone decomposition. Interdicip. Toxicol. 2014 ; Vol. 7(2) ; 47-59 doi : 10.2478/intox-2014-0008
[ 2] Daumont, D. et al.: Ozone UV spectroscopy I: Absorption cross-sections at room temperature., J. Atmos. Chem.15(2), 145–155 (1992). doi:10.1007/bf00053756
[3 ] Meyer, C. P. et al.: Water vapor interference in the measurement of ozone in ambient air by ultraviolet absorption. Rev. Sci. Instruments. 62(1), January 1991, 223-228.
Fig. 9: Determination of the detection limit as 3 times the standard deviation of the O3 measurements at zero point
[4 ] ISO 13964 (1998) Air quality- Determination of ozone in ambient air – Ultraviolet photometric method
[5 ] Wiegleb, G.: Einsatz von LED-Strahlungsquellen in Analysengeräten. Laser- und Optoelektronik No.3/1985, 308-310
[6] Wiegleb, G.: Gas Measurement Technology in Theory and Practice, Springer Verlag Wiesbaden 2023
[7] Merilo, E., et al.: PYR/RCAR Receptors Contribute to Ozone-, Reduced Air Humidity-, Darkness-, and CO2-Induced Stomatal Regulation. Plant Physiology, July 2013, Vol. 162, pp. 1652–1668,
Prof. Dr. Gerhard Wiegleb Wi.Tec-Sensorik GmbH
Fig. 10: Periodic exposure (15-minute intervals) of the structure to zero gas, followed by adjustment of the zero point. The variation in the fi nal value of ±10ppb O3
is due to the unstable generation of the O3 ozone generator. concentration in the
Schepersweg 41-61, D-46485 Wesel, Germany Email:
ge.wiegleb@witec-sensorik.de Web:
www.witec-sensorik.com
READ, SHARE or COMMENTon this article at:
envirotech-online.com/article
Page 1 |
Page 2 |
Page 3 |
Page 4 |
Page 5 |
Page 6 |
Page 7 |
Page 8 |
Page 9 |
Page 10 |
Page 11 |
Page 12 |
Page 13 |
Page 14 |
Page 15 |
Page 16 |
Page 17 |
Page 18 |
Page 19 |
Page 20 |
Page 21 |
Page 22 |
Page 23 |
Page 24 |
Page 25 |
Page 26 |
Page 27 |
Page 28 |
Page 29 |
Page 30 |
Page 31 |
Page 32 |
Page 33 |
Page 34 |
Page 35 |
Page 36