ReCap™ Total Resurfacing System
11d. Acetabular component insertion
Once the acetabular component has been
securely locked onto the impaction handle, the
ReCap acetabular component is impacted into the
prepared acetabulum at an angle of 45 degrees of
inclination and 20 degrees of anteversion. When
correct component seating has been achieved, the
impaction handle should be removed from the cup.
FLV HERE
This is achieved by turning the strike plate one-
quarter turn anti-clockwise. (Figure 25).
Should there be the need to carry out additional
impaction of the acetabular component, size
specific cup impactors are available on the ReCap
Figure 25
instrumentation tray (Figure 26).
To remove the strike plate from the impaction
handle, unscrew the strike plate until you fell
resistance. Lever the impactor plate from the
inserter handle with the multi-tool. (Figure 27).
FLV HERE
Figure 26
Figure 27
16