Understanding your P’s, Q’s, and S’s in instrument care & handling
by Stephen M. Kovach Q
ments available today in a medical device reprocessing department are made of stainless steel, which in most respects is an ideal material. It resists rust, takes a fi ne point, and retains an edge. But the term “stainless” is frequently, to me, taken too literally. Stainless steel does have its weaknesses—water spotting and staining are but two. So, there is no “stainless” type of steel.
A During manufacturing, surgical instru-
ments undergo many processes Two of which, we will focus, that help to inhibit the tendency to corrode are passivation and polishing. Passivation is a chemical process.
It should remove all iron particles, leaving a corrosion resistant-surface by forming a thin-transparent oxide fi lm. Passivation subjects the instrument to a solution of dilute acid and oxidizing salts. Steam sterilization helps in protecting the passivation layer and promoting its continued buildup. Also, the regular use of a water-soluble lubricant reinforces this protective coating. Rust and corrosion can occur if the passivation layer is damaged; thus, if the oxidized layer is destroyed or damaged, the instrument will have a greater tendency to rust. Polishing achieves
a smooth, closed- surface and builds up a protective outer layer of chro- mium oxide. This oxidized layer is highly resistant to corrosion and con- tinues to build up with regular use and sterilization.
I have been asked to, “Explain the difference between polishing and
passivation,” and, “Explain the differ- ence about spotting and staining on instruments.” I will tie both answers together because, in my mind, they go hand-in-hand when it comes to care and handling of surgical instruments.
In my observation, approximately 75% or more of the surgical instru-
Remember, on the other hand, this pro- tective coating could be unknowingly removed if the instruments are cleaned improperly (i.e., exposed to harsh clean- ing solutions and brushed with highly abrasive materials). Also, etching instru- ments will ruin the coating and promote corrosion. Another P, medical device reprocessing professionals should be aware of, is Pit- ting of the medical devices. Pitting can be caused by exposure to blood, chloride, or other highly corrosive solutions. It occurs in shallow to deep, localized defects that appear as black holes (pits) on the instru- ment’s surface. Some say you can’t repair this defect. What you must do is remove pitted instruments from the set and replace them, and work with your repair company to fi nd the causes.
Spotting is a discoloration in the stain- less steel of surgical instruments. Spotting differs from a stain in that it looks like it has been deposited on the surface rather than being a discoloration inherent in stainless steel. Staining is a discoloration of the stain- less-steel surface. Some stains disappear when the cause is removed or understood. Others may be permanent until the device is refurbished by the original manufac- turer of the device or an authorized dealer like a repair company. Some can never be refurbished, the damage is too far gone, once noticed. I have been talking about stainless-steel instruments, and, as I stated earlier, they represent at least 75% of a department’s instruments inventory. I want to stress that medical device reprocessing profes-
sionals need to understand all the various materials and how they are processed. One material is titanium. This is what one company states about the color on their instruments. “Different shades of blue may be found on our various anodized xxxx Titanium instrumentation. This is not a defect but a phenomenon, which can occur due to any slight variation of manu- facturing technique or alloy composition. This is a cosmetic characteristic only and does not compromise the high-quality of your xxxx Titanium Instrumentation.” This statement shows the importance of reviewing the instructions for use (IFU) of all devices.
Many device manufacturers have charts
within their instrument catalog explaining causes and solutions to common staining and spotting. If you use a repair company, they also have resources to help solve these issues.
My experience has shown me that many of the staining and spotting problems are due to such factors as a) water pH, b) boiler compounds in water lines, c) clean- ing solution residue, d) laundry residue from cloth towels/wrappers, e) exposure to chemicals incompatible with the surgi- cal instrument material, and f) automatic washers not being tested properly with clinically relevant evidence-based prod- ucts , to name a few of the reasons. Taking care of one of the medical facil- ity’s greatest investment medical devices used in surgery (surgical instruments), understanding these terms, and work- ing with your various manufacturers is important to make them last as long as possible. Staff needs to remember “inspect to prevent” at each stage of the medi- cal device repro- cessing cycle. And if an issue fi nd out the “why” to pre- vent it from taking place again. HPN
Figure 1: Dirty back-up instruments Figure 2: Dirty instruments with water spots
References online at ttps://hpnonline. com/21286602
hpnonline.com • HEALTHCARE PURCHASING NEWS • December 2022 41
Page 1 |
Page 2 |
Page 3 |
Page 4 |
Page 5 |
Page 6 |
Page 7 |
Page 8 |
Page 9 |
Page 10 |
Page 11 |
Page 12 |
Page 13 |
Page 14 |
Page 15 |
Page 16 |
Page 17 |
Page 18 |
Page 19 |
Page 20 |
Page 21 |
Page 22 |
Page 23 |
Page 24 |
Page 25 |
Page 26 |
Page 27 |
Page 28 |
Page 29 |
Page 30 |
Page 31 |
Page 32 |
Page 33 |
Page 34 |
Page 35 |
Page 36 |
Page 37 |
Page 38 |
Page 39 |
Page 40 |
Page 41 |
Page 42 |
Page 43 |
Page 44 |
Page 45 |
Page 46 |
Page 47 |
Page 48 |
Page 49 |
Page 50 |
Page 51 |
Page 52 |
Page 53 |
Page 54