What’s new in ANSI/AAMI ST91 latest update?
by Stephen M. Kovach
Q A
“I have heard that ANSI/AAMI ST91 has been approved. If that is true, what are some of the major changes we should be aware of?”
I have been told it is ready for distribution sometime this month. As you know, there are hundreds of changes to the commentary. Some are just editorial, while others are technical. Here’s my Top 10 list of things to know about the new ST91. 1. Precleaning renamed “Point of Use Treatment”: There is also now a recommendation to note the time when point of use treatment is conducted and convey that information to the processing staff so that they know whether or not to follow the delayed reprocessing procedure.
2. Leak testing procedure has been changed to 60 seconds of observation instead of 30 seconds: There is now a recommenda- tion to test the leak test units for pressure output every day the unit is used.
3. Added a designation of “high-risk” endoscopes: The standard contains diff erent guidance for processing high-risk versus non-high- risk endoscopes.
4. Cleaning verifi cation testing intervals have changed: ST91 now requires cleaning verifi cation on all high-risk endoscopes and a statistically signifi cant portion of non-high-risk endoscopes. The document walks you through how to fi gure that out in an annex.
5. Strengthened guidelines around enhanced visual inspection: The standard now recommends the use of lighted magnifi cation.
6. Borescope inspection is now recommended at a periodic interval determined by the facility.
7. Manual disinfection is no longer recommended due to the variability/inconsistency personnel responsible for the pro- cess: That’s a big one!
8. ST91 strengthened wording on which scopes should be steril- ized and which may be high-level disinfected: All endoscopes should be sterilized (if possible), but if not possible, then we resort to high-level disinfection. Endoscopes used in a critical application— meaning used in a sterile area of the body—must be sterilized.
9. All endoscopes are now required to be completely dried (both externally and internally) before reuse, even though they were run through an AER: This does not matter whether the endoscope is going into the storage cabinet or back to immediate reuse; all scopes should be subjected to a compressed air-drying cycle of 10 minutes or placed into a drying cabinet to dry internally. This will add time to the processing cycle.
10. Endoscopes should be handled with new, clean gloves (at all times): Perform hand hygiene before donning and after doffi ng gloves.
Photo credit: sukan |
stock.adobe.com
ST91 is a document that involved and re ects the collective expertise of a committee of healthcare professionals and indus- trial representatives. I want to thank the AAMI co-chairs: user co-chairs Nancy Chopin and her successor Garland Grisby and industry co-chair Mary Ann Drosnock (whose term just ended upon ST91’s comple- tion). Thank you for your hard work and dedication through the development of the new ST91 document.
As with any standard document (especially this one), my sug- gestion is to order your own copy from AAMI at www.aami. org to read the complete document. You asked for my thoughts about the changes in ASIAAI ST91. If you want an offi cial interpretation of any AAMI standards and recommended prac- tices, you must make it in writing to the AAMI Vice President, Standards Policy and Programs. Good luck in implementing ANSI/AAMI ST91 at your facility. HPN
hpnonline.com • HEALTHCARE PURCHASING NEWS • January 2022 39
Page 1 |
Page 2 |
Page 3 |
Page 4 |
Page 5 |
Page 6 |
Page 7 |
Page 8 |
Page 9 |
Page 10 |
Page 11 |
Page 12 |
Page 13 |
Page 14 |
Page 15 |
Page 16 |
Page 17 |
Page 18 |
Page 19 |
Page 20 |
Page 21 |
Page 22 |
Page 23 |
Page 24 |
Page 25 |
Page 26 |
Page 27 |
Page 28 |
Page 29 |
Page 30 |
Page 31 |
Page 32 |
Page 33 |
Page 34 |
Page 35 |
Page 36 |
Page 37 |
Page 38 |
Page 39 |
Page 40 |
Page 41 |
Page 42 |
Page 43 |
Page 44 |
Page 45 |
Page 46 |
Page 47 |
Page 48 |
Page 49 |
Page 50 |
Page 51 |
Page 52