Technical Specification Measuring principle


Combustor operating


temperature range Heating rate Sample size


Detection limit Repeatability Dimensions


Pyrolysis combustion flow calorimetry 25 to 1000°C


0.4 to 4°C/s 1 to 5mg 5mW


± 5% (1 mg sample) 1050mm (H) x 350mm (W) x 550mm (D)


Built in accordance with EMC 89/336/EEC, LVD 72/23/EEC, BS EN 60204-1, BS EN 746-2 Due to the continuous development policy of FTT technical changes could be made without prior notice.


Services Gas supply


Power Extraction Scales Drierite


Pressure regulated Nitrogen (oxygen free) at approximately 2 bar Pressure regulated Oxygen at approximately 2 bar


Dual voltage 96-264VAC 10A, 50/60Hz


Fume cupboard or extraction system capable of extracting a volume flow of 100cc/min


With a capacity of 250 mg or greater and a sensitivity of 0.01mg, to weigh specimens or containers, or both.


Drierite is required to remove the water from the sample.


Charlwoods Road East GrinsteadWest Sussex RH19 2HL United Kingdom Tel: +44 (0)1342 323600 Fax: +44 (0)1342 323608 email: sales@fire-testing.com web: www.fire-testing.com