Technical Specification Measuring Principle


Heat Flux Range Specimen Dimensions Instrument Dimensions (approximate) Services


Test Room –The FRP apparatus should be situated in a draught free environment at 23 ± 5°C and a relative humidity of 50 ± 20%. Electrical Supply – 230VAC 50/60Hz, 12 Amps Hood –The apparatus should be situated under a suitable extraction system.


Gas – Propane at a pressure of 2 bar with a calorific value of approximately 83MJ/m3 is required for the radiant panel and the pilot burner. Compressed air at a pressure of 6-7 bar is required for the radiant panel.


Water –Water at 15-25°C is required for cooling the total heat flux meter.


Surface spread of flame on flooring material exposed to radiant heat source


1.0 kW/m2 – 10.9 kW/m2 1050mm (L) x 230mm (W) 1900mm (W) x 750mm (D) x 1900mm (H) Due to the continuous development policy of FTT technical changes could be made without prior notice.